Belcat II

Belcat II

microtrac-bel

Building on the experience and technology of the BELCAT series, the all-new fully-automated catalyst analyzer BELCAT II has been developed.

Download: brochure_bel_belcat-II_en
Link

Product Info

  • Details
  • Specifications

BELCAT II is the analyzer of choice for catalyst researchers and enables comprehensive catalyst evaluation by using the following techniques:

  • Pulse chemisorption
  • Temperature programmed desorption (TPD)
  • Temperature programmed reduction (TPR)
  • Temperature programmed oxidation (TPO)
  • Temperature programmed reaction (TPReaction)
  • BET single point measurement

TYPICAL APPLICATIONS

  • Acidity study of cracking catalyst by ammonia or pyridine TPD
  • Basicity study of solid base catalyst
  • Metal dispersion rate evaluation of exhaust gas catalyst / fuel cell electrode catalyst
  • Determination of optimal preparation conditions of catalyst
  • Catalytic reaction study

 

PRODUCT ADVANTAGES

  • Affordable price
  • Small foot print
  • Precise gas control by three high performance mass flow controllers
  • Precise automatic gas injection system for calibration
  • The pulse loop with temperature/pressure sensor enables the highly reproducible result
  • A variety of options to fulfill the customers' demand
  • With the standard gas mixing equipment, mixed gases (H2/Ar, O2/He, etc...) are not necessary

TECHNICAL DATA

Measurement principle Dynamic flow method
Detector Semi-diffusion type 4-element thermal conductivity detector (TCD)
Measurement/pretreatment port 1
Injection gas He, Ar, N2, O2, H2, CO, CO2, NH3, N2O, NO, etc...
Gas Carrier: 3
Pretreatment/Pulse/MIX: 8 (Corrosion-resistant x 2)
Mass Flow Controller Carrier: F.S.100sccm
Pretreatment/Pulse: F.S.100sccm
MIX: F.S.30sccm
Electric furnace Regular temperature: 1,100°C
Maximum temperature: 1,200°C
Quick cooling: 30min (400°C->50°)
CATCryo II (option): -120°C supported
Vapor injection (option) H2O, CH3OH, C2H5OH, toluene, benzene, etc...
Dimensions (W x H x D) ~ 500 x 750 x 500 mm
Weight ~ 80 kg
Utility: Gas Measurement gas: 0.1MPa(G)
Valve driving: 0.45~0.55MPa(G)
J1/8"/8h" Swagelok connection
Utility: Power AC110V/220V, 1,300W

Content may be subject to modifications or corrections